Thermo Scientific TVA 2020

Daily Rate $180

Weekly Rate $500

Monthly Rate $1490

Quickly detect fugitive emissions of organic and inorganic compounds for Method 21 compliance, LDAR applications, and site remediation with this lightweight, intrinsically safe portable gas leak detector. The Thermo Scientific™ TVA2020 Toxic Vapor Analyzer can be configured as an FID analyzer (flame ionization detector) or dual detection FID analyzer and PID analyzer (photoionization detector). Key features include reduced response times, increased user serviceability, and Bluetooth connectivity.

Description


TVA 2020

Toxic Vapor Analyzer

The TVA2020 Toxic Vapor Analyzer is 21% lighter than earlier models and more compact than most stand alone FID analyzers. The lightweight and compact design reduces fatigue for true portability and is easier to service in the field. In addition, a variety of options are available such as a basic or enhanced probe, carrying case, and hydrogen refill assembly.
Optional Bluetooth communication transmits concentration data to a handheld device containing the LDAR software database thereby eliminating the need to transfer files post-monitoring and provide a greater access to route information


The Thermo Scientific TVA2020 Toxic
Vapor Analyzer is capable of detecting
virtually all organic and inorganic
compounds. The TVA2020 analyzer
can be configured for use in diverse
applications including U.S. EPA Method
21 monitoring, site remediation, landfill
monitoring, and general area surveys.
The TVA2020 analyzer is equipped
with a Flame Ionization Detector to
measure organic compounds with high
sensitivity.

The FID technology allows
for a wide dynamic and linear range
that produces stable and repeatable
responses. The analyzer can be
configured with both FID and PID
technology for simultaneous detection
and enhanced analytical capabilities.
This dual configuration is capable of
producing a more rapid reading of organic
and inorganic compounds as opposed
to a single detector technology and
provides more comprehensive gas coverage than comparable size devices.
After performing a primary calibration,
the TVA2020 analyzer can be customized
by activating internal logging parameters,
uploading a monitoring route, establishing a bluetooth connection, setting alarm
levels, and activating response factors.


Features

Easy to Use

The TVA2020 is 21% lighter than earlier models and more compact than most stand alone FID analyzers. The lightweight and compact design reduces fatigue for true portability and is easier to service in the field.

Efficiently Made

Detect gas leaks from refineries and chemical plants to comply with U.S. EPA Method 21 and LDAR programs. The TVA2020 Toxic Vapor Analyzer is equipped with an FID to measure organic compounds with high sensitivity.

Bluetooth

Detect gas leaks from refineries and chemical plants to comply with U.S. EPA Method 21 and LDAR programs. The TVA2020 Toxic Vapor Analyzer is equipped with an FID to measure organic compounds with high sensitivity.

  • Real-time gas detector readings and alarm status enabled by state of the art wireless technology

  • Unmistakable five-way local and remote wireless notification of alarm conditions

  • Intuitive simple-to-operate two-button user interface

  • Strong, protective, concussion-proof design

  • IP-65/67 water- and dust-resistant case

  • Easy maintenance with fiel dreplaceable sensors and pump

  • Pumped or diffusion models available

Specifications

Accuracy
FID Instrument
±10% of reading or ±1.0 ppm, whichever is greater,
from 1.0 to 10,000 ppm.
PID Instrument
±20 % of reading or ±0.5 ppm, whichever
is greater, from 0.5 to 500 ppm.
Repeatability
Fid Instrument
±2% at 500 ppm of methane
PID Instrument
±1% at 100 ppm of isobutylene
Linear Range
Fid Instrument
1.0 to 50,000 ppm of methane
PID Instrument
0.5 to 2,000 ppm of isobutylene
Relative Humidity Range
15 – 95%


Response Time
Fid Instrument
Less than 3.5 seconds for 90% of final value, using
10,000 ppm of methane
Pid Instrument
Less than 3.5 seconds for 90% of final value, using
500 ppm of isobutylene
Sample Flow Rate
1 liter/minute, nominal, at sample probe inlet
Hydrogen Supply Operating Time
10 hours of continuous operation, starting from a cylinder charged
up to 15.3 MPa (2200 psi)
Physical Dimensions
11.5" × 9" × 4" (29.2 cm × 22.9 cm × 10.2 cm)
Weight
Fid only
9.2 lbs
Dual
9.4 lbs


Minimum Detectable Limit
The minimum detectable level is defined as seven times
the standard deviation of peak-to-peak noise.
Fid Instrument
0.5 ppm of methane
Pid Instrument
0.5 ppm of isobutylene
Battery
The battery operating time is 10 hours minimum at 0 °C (32 °F).
Fully charged in less than 10 hours.
Lamp File
Fid Instrument
Greater than 5,000 hours
Pid Instrument
Greater than 2,000 hours, with normal cleaning
Data Storage Interval
Auto mode
1 per second to 1 per 999 minutes, user selectable
VOC or FE Mode
2 to 30 seconds, user selectable

Resources

Media

Thermo Scientific TVA2020 Toxic Vapor Analyzer

Thermo Scientific Model TVA2020 Toxic Vapor Analyzer